Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 255
![background image](/manuals/33909/255/background.png)
255
Specifications
Performance Specifications
Analog output
For pressure monitoring, 2 mV/bar, one output
Communications
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start,
stop and shut-down signals, LAN optional
Safety and maintenance
Extensive diagnostics, error detection and display (through control module
and Agilent ChemStation), leak detection, safe leak handling, leak output
signal for shutdown of pumping system. Low voltages in major
maintenance areas.
GLP features
Early maintenance feedback (EMF) for continuous tracking of instrument
usage in terms of seal wear and volume of pumped mobile phase with
user-settable limits and feedback messages. Electronic records of
maintenance and errors.
Housing
All materials recyclable.
Table 44
Performance Specification Agilent 1100 Series Quaternary Pump