Agilent 1100 to 3392/3 integrators, Agilent 1100 to 3394 integrators – Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 186
![background image](/manuals/33909/186/background.png)
186
Parts and Materials
Remote Cables
Agilent 1100 to 3392/3 Integrators
Agilent 1100 to 3394 Integrators
N O T E
START and STOP are connected via diodes to pin 3 of the 3394 connector.
Connector
01046-60206
Pin
3392/3
Pin
Agilent 1100
Signal Name
Active
(TTL)
3
1 - White
Digital ground
NC
2 - Brown
Prepare run
Low
11
3 - Gray
Start
Low
NC
4 - Blue
Shut down
Low
NC
5 - Pink
Not connected
NC
6 - Yellow
Power on
High
9
7 - Red
Ready
High
1
8 - Green
Stop
Low
NC
9 - Black
Start request
Low
4 - Key
Connector
01046-60210
Pin
3394
Pin
Agilent 1100
Signal Name
Active
(TTL)
9
1 - White
Digital ground
NC
2 - Brown
Prepare run
Low
3
3 - Gray
Start
Low
NC
4 - Blue
Shut down
Low
NC
5 - Pink
Not connected
NC
6 - Yellow
Power on
High
5,14
7 - Red
Ready
High
6
8 - Green
Stop
Low
1
9 - Black
Start request
Low
13, 15
Not connected