Agilent Technologies Quaternary Pump 1100 Series User Manual
Page 139
![background image](/manuals/33909/139/background.png)
139
Repairing the Pump
Exchanging the Damper
3
Disconnect the two damper capillaries.
4
Loosen the screws of the Z-panel and take it
out of the instrument.
5
Disconnect the damper connector at the main
board (J24). Lift the damper out of its
position.
6
Place the new damper into the recess and
connect to the main board (J24).
Damper
capillaries
Screws
Z-panel
Damper
Damper
Connector
Damper
Connector
Damper