2610 DO Sensor

3-2610-31 to 9900 Set Up

1. Wire the 9900, 2610-31, and 8058 as shown in the diagrams.
2. On the 9900 press and

hold the ENTER key for 2
seconds.

3. Press

the

▼ key to select the

INPUT menu item.

4. Press the ENTER key to

access the INPUT menus.

5. Press

the▲ key to select the

TYPE menu item.

6. Press

the

► key to change

the TYPE selection.

7. Enter the password code.
8. Repeatedly press the ▼ key

until the “4-20 mA INPUT”
menu item is fl ashing and
then press the ENTER key.

9. The 9900 will prompt “All

settings will be reset”, press the
▲ key to select “Yes” and then press the ENTER key.

10. The 9900 will return to the View Mode display.
11. Press and hold the ENTER key for 2 seconds.
12. Press the ▼ key to select the INPUT menu item.
13. In the NAME menu item, press the ► key to change the

displayed name from “4-20 mA INPUT” to a more descriptive
name (e.g., "DISSOLVED O2") and press ENTER when done.

14. Press the ▼ key to select SENSOR UNIT menu item.
15. Press the ► to change the label from UNIT to MG/L or PPM

and press ENTER.

16. Press the ▼ key and ensure the 4 mA VALUE is set to 0.0000.
17. Press the ▼ key and change the 20 mA VALUE from

5.0000 to 20.000 and press ENTER.

18. Press both the▲ and ▼ keys simultaneously to return to

the Menu.

19. Press the ▼ to select the LOOP menu and press ENTER.
20. Set the 4 MA SET POINT to your desired value. The 2610 is

factory set for a 0 to 20 mg/L output. Press ENTER when done.

21. Press the ▼ key to select the 20 MA SET POINT and set to your desired value. The 2610 is factory set for a

0 to 20 mg/L output. Press ENTER when done.

22. Press both the ▲ and ▼ keys simultaneously to return to the Menu.
23. Press the ▼ key twice to select the OPTION menu and press ENTER.
24. Press the ▼ key twice to select the SET BAR MIN option. Change this option if desired. The 2610 is factory

set for a 0 to 20 mg/L output. Press the ENTER key when done.

25. Press the ▼ key to select the SET BAR MAX option. Change this option if desired. The 2610 is factory set for

a 0 to 20 mg/L output. Press ENTER when done.

26. Press both the▲ and ▼ keys simultaneously to return to the Menu.
27. ENTER the other menus and set the unit as desired for your application.
28. Press both the▲ and ▼ keys simultaneously to return to the View Menu.

RED

DO Sensor

3-2610-31

24 VDC

Regulated

Power Supply

BLACK

Input

4-20 mA

Output

S

3

L

SHIELD

(no connection)

Signet 8058 i-Go™

4-20 mA to S

3

L Converter

+GF+

4-20 mA input S

3

L Output

V–

V+

BROWN

GREEN

BLUE

WHITE

No connection

8058-1

PWR–

RED

WHITE

SHIELD

BLACK

DA
T

A

GND

SHLD

V+

9900 S

3

L Inputs

9900 Power Inputs

LOOP–

LOOP+

PWR+

PWR–

LOOP–

LOOP+

PWR+

DA
T

A

GND

V+

SHLD

+

-

3 2 1

6 5 4

7 8 9

Loop1

PWR

N/C

S L

DATA

3

Loop2

PWR

+GF+ SIGNET

BROWN

24 VDC

Regulated

Power Supply

DO Sensor 1

3-2610-31

No connection

9900 Power Inputs

9900 S

3

L Inputs

8058-2

BLACK (5 VDC)

BLACK

BLACK

RED

RED

WHITE (GND)

RED (S

3

L)

Technical Notes:
• The cable length from the 8058 to the 9900 must not exceed 60 m (200 ft).
• When using the 8058-2 with the 9900, connect the loop source to 8058-2 Channel 1 input

ONLY as shown in the fi gure.

• See the 8058 manual for more information.

Wiring and Set Up