Fig. 33 – Burkert Type 8041 User Manual
Page 40
38
Adjustmentandcommissioning
Calibration in progress
→
Press and hold the push
button until...
to select the calibration of the
other parameter
B
A
→
Briefly press push button.
End of calibration: red LED
goes off?
Go back to Read mode
.....
NO
to calibrate automatically
Red LED flashes:
4 times: fluid velocity < 2 m/s
5 times: fluid velocity < 10 m/s
The full scale cannot be calibrated
→
Press push button for 2 s.
Go back to Read mode using the
full scale selected or previously
calibrated.
Full scale value stored
YES
→
Wait for 10 s.
to go back to the read mode
C
Go back to Read mode without con-
firming the selection displayed.
...bargraph goes back to step
1, Fig. 32
Fig. 33: Calibration of the full scale, part 2
English
Type 8041
See also other documents in the category Burkert Sensors:
- Type 1062 (112 pages)
- Type 1050 (4 pages)
- Type 8750 (64 pages)
- Type 8750 (82 pages)
- Type 8681 (40 pages)
- Type 8681 (90 pages)
- Type 8798 (2 pages)
- Type 8791 (4 pages)
- Type 8798 (106 pages)
- Type 8791 (15 pages)
- Type 8792 (136 pages)
- Type 8792 (252 pages)
- Type 8718 (34 pages)
- Type 8792 (118 pages)
- Type 8791 (106 pages)
- Type 8791 (184 pages)
- Type 8791 (28 pages)
- Type 8791 (21 pages)
- Type 8791 (154 pages)
- Type 0911 (64 pages)
- Type 0911 (84 pages)
- Type 0911 (76 pages)
- Type 0911 (46 pages)
- Type 1058 (31 pages)
- Type 1060 (4 pages)
- Type 1066 (112 pages)
- Type 1067 (158 pages)
- Type 1077-2 (33 pages)
- Type 1094 (82 pages)
- Type 1094 (126 pages)
- Type 1094 (12 pages)
- Type 1094 (41 pages)
- Type 1115 (25 pages)
- Type 1150 (99 pages)
- Type 1541 (2 pages)
- Type 5142 (6 pages)
- Type 8619 (40 pages)
- Type 8619 (134 pages)
- Type 8620 (177 pages)
- Type 8622 (4 pages)
- Type 8623 (90 pages)
- Type 8623 (130 pages)
- Type 8625 (118 pages)
- Type 8624 (124 pages)
- Type 8718 (4 pages)