INstallation
Vertical mounting
Correct
Incorrect
flow direction
Fig. 10: Filling of the pipe
Correct
flow direction
Incorrect
Correct
Incorrect
Fig. 11: Air bubbles within the pipe
→
If necessary, use a flow conditioner to improve measurement
precision.
8.2.2. installation into the pipe of a 8041
with a G2'' nut
Observe the installation recommendations described at
chap. 8.2.1 and in the operating instructions of the S020.
→
Install the S020 fitting on the pipe.
→
Insert the nut (see mark 3, Fig. 12) on the fitting.
→
Insert the snap ring (mark 2, Fig. 12) into the groove (mark 5,
Fig. 12).
→
Position the cable glands parallel to the pipe and insert the
device (mark 1, Fig. 12) into the fitting.
→
Tighten the nut (mark 3, Fig. 12) by hand on the device.
1
2
3
5
4
Electrodes
Flow direction
Fig. 12: Installation into the pipe of the flowmeter with a G2'' nut
English
Type 8041
- Type 1062 (112 pages)
- Type 1050 (4 pages)
- Type 8750 (64 pages)
- Type 8750 (82 pages)
- Type 8681 (40 pages)
- Type 8681 (90 pages)
- Type 8798 (2 pages)
- Type 8791 (4 pages)
- Type 8798 (106 pages)
- Type 8791 (15 pages)
- Type 8792 (136 pages)
- Type 8792 (252 pages)
- Type 8718 (34 pages)
- Type 8792 (118 pages)
- Type 8791 (106 pages)
- Type 8791 (184 pages)
- Type 8791 (28 pages)
- Type 8791 (21 pages)
- Type 8791 (154 pages)
- Type 0911 (64 pages)
- Type 0911 (84 pages)
- Type 0911 (76 pages)
- Type 0911 (46 pages)
- Type 1058 (31 pages)
- Type 1060 (4 pages)
- Type 1066 (112 pages)
- Type 1067 (158 pages)
- Type 1077-2 (33 pages)
- Type 1094 (82 pages)
- Type 1094 (126 pages)
- Type 1094 (12 pages)
- Type 1094 (41 pages)
- Type 1115 (25 pages)
- Type 1150 (99 pages)
- Type 1541 (2 pages)
- Type 5142 (6 pages)
- Type 8619 (40 pages)
- Type 8619 (134 pages)
- Type 8620 (177 pages)
- Type 8622 (4 pages)
- Type 8623 (90 pages)
- Type 8623 (130 pages)
- Type 8625 (118 pages)
- Type 8624 (124 pages)
- Type 8718 (4 pages)