4 model 5866 orifice selection -14, Reactor n, Model 5866 vent to atmosphere h – Brooks Instrument 5866M User Manual

Page 37: 100 psig maximum desired pressure

background image

Brooks Instrument Model 5866 Pressure Controller

5-14

Since the C

v

calculated is the same as listed for the .014" orifice the .020"

orifice should be selected. The last step is to check the maximum pressure
drop for the application. The maximum pressure drop for the .020 orifice is
300 psi, this is greater than the 100 psi pressure drop required for the
application so the .020" orifice is correct.

Table 5-4. Model 5866 Orifice Selection.

Orifice

Maximum

C

v

Size

Pressure Drop (psi)

5.0x10

-6

.001

1500

2.0x10

-5

.002

1500

5.0x10

-5

.003

1500

1.1x10

-4

.004

1500

4.0.x10

-4

.007

1500

9.0x10

-4

.010

750

2.0x10

-3

.014

550

4.7x10

-3

.020

300

1.4x10

-2

.032

175

3.5x10

-2

.048

85

5.5x10

-2

.062

55

8.0x10

-2

.078

50

1.0x10

-1

.093

50

1.4x10

-1

.116

50

1.5x10

-1

.120

50

Figure 5-6. Example Application of Orifice Sizing Procedure.

®

Reactor

N

2

Model 5850E
Mass Flow Controller
500 sccm, Full Scale

®

Model 5866

Vent to

Atmosphere

H

2

Model 5850E
Mass Flow Controller
300 sccm, Full Scale

®

100 psig

Maximum

Desired

Pressure