beautypg.com

1 d-connector pin arrangement -5, Ansducer signal (refer to section 2-11), Table 2-1 – Brooks Instrument 5866M User Manual

Page 16: 9 valve override, 10 precision 5v reference, 11 remote transducer input, 12 valve test point

background image

2-5

Installation

Table 2-1. D-Connector Pin Arrangement.

PIN NO.

FUNCTION

COLOR CODE

1

Command Common (Command Pot "CCW")

Black

2

Voltage Signal Output

White

3

Remote Valve Return

Red

4

Current Output

Green

5

+15 Vdc Supply

Orange

6

-15 Vdc Supply

Blue

7

Valve Voltage Out/External Valve Control

Wht/Blk

8

Command Input (Command Pot "S")

Red/Blk

9

Supply Voltage Common

Grn/Blk

10

Voltage Signal Common

Org/Blk

11

+5 Volt Reference (Command Pot "CW")

Blu/Blk

12

Valve Override

Blk/Wht

13

+15 Vdc Supply (1)

Red/Wht

14

Chassis Ground (2)

Grn/Wht

15

External Sensor Input

Blu/Wht

NOTES:
1. Pin 13 does not have to be

connected if current I/O is
not used.

2. Cable shield tied to chassis

ground in meter connector.

1

15

9

8

2-9 Valve Override

The valve override function allows full opening and closing of the valve
independent of the command setting.

The valve override for the pressure controller is as follows:
a. To open the valve, apply +15 Vdc to the valve override terminal.
b. To close the valve, apply -15 Vdc to the valve override terminal.

The valve override function is accessed from Terminal 12 (refer to Figure
2-1 and
Table 2-1).

Note: For normal operation, Terminal 12 may be left open (floating) or
grounded.

2-10 Precision 5V Reference

The Model 5866 is equipped with a precision 5 Vdc reference. The
reference is used internally for precise zero adjustment and is also
available as a buffered output for use in conjunction with a 1 to 10 k ohm
potentiometer to provide 0-5 Vdc set-point. The 5 Vdc reference is
available at Terminal 11 (refer to Figure 2-1 and Table 2-1).

2-11 Remote Transducer Input

To allow the Model 5866 control circuitry and valve to be used with an
external transducer signal such as a capacitance manometer, Jumper J2
“C” or “D” must be moved to the Position B. Jumper J2, "A" can be in either
Position a or b depending on valve/controller type. The external sensor
signal (0-5 or 0-10 Vdc) must be connected to Terminals 15 and 10 (refer
to Figure 2-1 and Table 2-1).

2-12 Valve Test Point

The valve voltage may be monitored on Terminal 7 of the D-Connector.

This manual is related to the following products: