Part 4 - software utilities – Dynasonics MFX Series Insertion MagProbe Flow Meter User Manual
Page 66

Rev. 01/11
-4.14-
MFX
5. Enter a value in the Set -> box that is equal to:
New Set -> = (Forward reading - Reverse reading)/2 + Set ->
6. Enter the new calculated Set -> value into the box, ensuring to
use a value of proper polarity—either positive or negative.
7. Press Next and then Finish on Page 3 of 3.
Saving Meter Configuration on a PC
The complete configuration of the flow meter can be saved from the
Configuration screen. Select File Save button located in the lower
left-hand corner of the screen and name the file. Files are saved as
a *.dcf extension. This file may be transferred to other flow meters or
may be recalled should the same pipe be surveyed again or multiple
meters programmed with the same information.
Printing a Flow Meter Configuration and Calibration Report
Select File from the upper task bar and Print to print a calibration/
configuration information sheet for the flow meter installation.
PART 4 - SOFTWARE UTILITIES
Wait for Stable Reading
Figure 4.11
Setting Zero Flow