x MAX Flow is used by the MFX to establish filter settings in its

operating system. Enter a flow rate that is the maximum, positive
flow rate anticipated within the system.

x The Damping value is increased to increase stability of the flow

rate readings. Damping values are decreased to allow the flow
meter to react faster to changing flow rates.

x Low Flow Cutoff is entered as a percentage between MAX

Flow and MIN Flow and influences how the flow meter will act at
flows very close to zero. Generally, an entry of 1% provides for a
stable zero indication, while providing a 100:1 turndown ratio for
measurements.

x Low Signal Cutoff does not pertain to the MFX product.
x Substitute Flow is used to provide an indication and output that

signifies that an error exists with the flow meter or its setup. It is
set as a percentage between MIN Flow and MAX Flow. In a
unidirectional system this value is typically set to zero, to
indicate zero flow while in an error condition. In a bi-directional
system, the percentage can be set such that zero is displayed in
an error condition. To calculate where to set the Substitute Flow
value in a bi-directional system, perform the following operation:

PART 4 - SOFTWARE UTILITIES

Exponent

Display Multiplier

E-1

0.1 (decimal is moved on display)

E0

1 (no multiplier)

E1

10

E2

100

E3

1,000

E4

10,000

E5

100,000

E6

1,000,000

TABLE 4.1

Totalizer Exponent Values

(

)

Substitute Flow = 100

100

× MAX Flow

MAX Flow

— MIN

Flow