beautypg.com

Part 3 - keypad configuration, Engineering units—total exponent – Dynasonics MFX Series Insertion MagProbe Flow Meter User Manual

Page 40

background image


Rev. 01/11

-3.6-

MFX

TOTL E -- Flow Totalizer Exponent Value

E-1 to E6


Utilized for setting the flow totalizer exponent. This feature is useful
for accommodating a very large accumulated flow. The exponent
is a “

10

n

” multiplier, where “n” can be from -1 (

0.1) to +6

(

1,000,000).Table 3.5 should be referenced for valid entries

and their influence on the DMFX display.

Table 3.5 - Totalizer Exponent Values

MIN RATE -- Minimum Flow Rate Settings (Value)

Rate Unit/Rate Interval

A minimum volumetric flow rate setting is entered to establish filter
software settings. Volumetric entries will be in the Engineering Rate
Units and Interval selected on pages 3.9 and 3.10 of this manual.
For unidirectional measurements, set MIN RATE to zero. For bi-
directional measurements, set MIN RATE to the highest negative
(reverse) flow rate expected in the piping system.


MAX RATE -- Maximum Flow Rate Settings (Value)

Rate Unit/Rate Interval

A maximum volumetric flow rate setting is entered to establish filter
software settings. Volumetric entries will be in the Engineering Rate
Units and Interval selected on pages 3.9 and 3.10 of this manual.
Set MAX RATE to the highest (positive) flow rate expected in the
piping system.

PART 3 - KEYPAD CONFIGURATION

Engineering
Units—TOTAL
Exponent

Exponent

Display Multiplier

E-1

0.1 (÷10)

E0

1 (no multiplier)

E1

10

E2

100

E3

1,000

E4

10,000

E5

100,000

E6

1,000,000

Minimum Flow
Rate

Maximum Flow
Rate