Part 4 - software utilities – Dynasonics MFX Series Insertion MagProbe Flow Meter User Manual
Page 55

Rev. 01/11
-4.3-
MFX
Click on the button labeled Configuration for updating flow range,
liquid, pipe and I/O operating information. The first screen that
appears after clicking the Configuration button is the BASIC tab.
See Figure 4.2.
1. BASIC TAB - see Figure 4.2
x General Units allows selection of either English (U.S.) or Metric
units of measure. If measurements of the pipe are to be entered
in inches, select English. If pipe measurements are to be entered
in millimeters, select Metric. If the General Units are altered from
those at instrument startup, then click on the Download button
on the lower right-hand portion of the screen and recycle power
to the MFX.
x Transducer Type does not pertain to the MFX product.
x Transducer Mount does not pertain to the MFX product.
x Transducer Spacing does not pertain to the MFX product.
x Pipe Material does not pertain to the MFX product.
x Pipe O.D. and Wall Thickness are based on the physical
dimensions of the pipe in which the probe will be inserted. Enter
this value in inches for English units or millimeters for Metric units.
x Liner Material is selected from the pull-down list.
Pipe and
Liquid
Configuration
Figure 4.2
Basic Tab
PART 4 - SOFTWARE UTILITIES