Range
From
0 to 6BD
Default
6BD
Related
Parameters
A, B, D, I, and M
Example 1
M=0
S=0 [t
L
= 3 μs]
D=0 (not for Bonito CL-400B/C 200fps cameras)
N (see table on next page)
Example 2
M=0
S=1 [t
L
= 1.5 μs]
D=0 (not for Bonito CL-400B/C 200fps cameras)
N (see table below)
IOD
IOD=0 if Continuous mode (e.g.
M=0),
IOD=1 if not in Continuous mode
t
L
t
L
= 3 μs in single CL channel mode (e.g.
S=0),
t
L
= 1.5 μsin dual CL channel mode (e.g.
S=1)
Setting
No. of lines
Minimum frame duration t
FDmin
Frame rate/fps
N=6BD
1726
5.181 ms
193.01
N=681
1666
5.001 ms
199.96
N=14B
332
0.999 ms
1001.0
N=1F
32
0.099 ms
10101
N=0
1
0.006 ms
166667
Table 20: Timing examples (M=0, S=0, D=0)
Setting
No. of lines
Minimum frame duration t
FDmin
Frame rate/fps
N=6BD
1726
2.5905 ms
386.03
N=681
1666
2.5005 ms
399.92
N=14B
332
0.4995 ms
2002.0
N=1F
32
0.0495 ms
20202
N=0
1
0.0030 ms
333333
Table 21: Timing examples (M=0, S=1, D=0)