Table 1-6

Uncorrected system performance (correction: off, system correction: on)

Description

Specification

3 MHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 8.5 GHz

Directivity

25 dB

20 dB

15 dB

Source match

25 dB

20 dB

15 dB

Load match

17 dB

12 dB

10 dB

Transmission tracking

±1.0 dB

±1.0 dB

±1.0 dB

Reflection tracking

±1.0 dB

±1.0 dB

±1.0 dB

Table 1-7

Test port output frequency

Description Specification

Supplemental

information

Range

E5070B

300 kHz to 3 GHz

E5071B

300 kHz to 8.5 GHz

Resolution 1

Hz

Source stability

Option E5070B/E5071B-UNQ

±5 ppm (5° C to 40° C, typical)

Option E5070B/E5071B-1E5

±0.05 ppm (23° C ± 5° C, typical)

±0.5 ppm/year (typical)

CW accuracy

Option E5070B/E5071B-UNQ

±5 ppm, 23° C ± 5° C

Option E5070B/E5071B-1E5

±1 ppm, 23° C ± 5° C

Uncorrected system performance

Test port output (source)