beautypg.com

E5092a configurable multiport test set – Atec Agilent-E5071B User Manual

Page 31

background image

31

E5092A configurable multiport test set

The section provides test input/output performance without calibration by the E5071B.

Table 3-1

Test set input/output performance

Description Specification Typical

Frequency range

50 MHz to 20 GHz

Damage level

20 dBm, ±35 VDC

Table 3-2

Option E5092A-020 port performance

Description Specification Typical

Load match (selected port)

SPDT switch

1

50 MHz to 2 GHz

17 dB

2 GHz to 4 GHz

11 dB

4 GHz to 8 GHz

8 dB

8 GHz to 10 GHz

7 dB

10 GHz to 18 GHz

4 dB

18 GHz to 20 GHz

4 dB

SP4T switch

2

50 MHz to 2 GHz

17 dB

2 GHz to 3 GHz

11 dB

3 GHz to 8 GHz

8 dB

8 GHz to 10 GHz

7 dB

10 GHz to 18 GHz

4 dB

18 GHz to 20 GHz

4 dB

Load match (unselected port)

SPDT switch

1

50 MHz to 3 GHz

17 dB

3 GHz to 10 GHz

11 dB

10 GHz to 16 GHz

8 dB

16 GHz to 18 GHz

6 dB

18 GHz to 20 GHz

4 dB

SP4T switch

2

50 MHz to 3 GHz

17 dB

3 GHz to 10 GHz

11 dB

10 GHz to 16 GHz

8 dB

16 GHz to 18 GHz

6 dB

18 GHz to 20 GHz

4 dB

1

SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 3-5.)