beautypg.com

Atec Agilent-E5071B User Manual

Page 4

background image

4

Table 1-2 Corrected system performance with type-N device connectors, 85032F calibration kit

Network analyzer: E5070B/E5071B, calibration kit: 85032F (type-N, 50 Ω), calibration: full 2-port

IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature = 23°C ±5°C with < 1°C deviation from calibration
temperature, isolation calibration not omitted

Description

Specification

(dB)

3 MHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 8.5 GHz

Directivity 49 40 38

Source match

41

36

35

Load match

49

40

37

Reflection tracking

±0.011

±0.032

±0.054

Transmission tracking

±0.016

±0.062

±0.088

Transmission uncertainty (specification)

Magnitude Phase

Reflection uncertainty (specification)
Magnitude Phase