Table 1-5 Corrected system performance with 3.5 mm device connector type, 85093C electronic calibration module

Network analyzer: E5070B/E5071B, calibration module: 85093C (3.5 mm, 50 Ω) electronic calibration (ECal) module, calibration: full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23° C ± 5° C with < 1° C deviation from
calibration temperature, isolation calibration not omitted

Description

Specification

(dB)

3 MHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 8.5 GHz

Directivity 52 51 47

Source match

44

39

34

Load match

47

44

40

Reflection tracking

±0.030

±0.050

±0.070

Transmission tracking

±0.039

±0.069

±0.117

Transmission uncertainty (specification)

Magnitude Phase

Reflection uncertainty (specification)
Magnitude Phase