10

Table 1-9

Test port output signal purity

Description Specification

Supplemental

information

Harmonics (2nd or 3rd)

10 MHz to 2 GHz

< -25 dBc (at 5 dBm, typical)

2 GHz to 3 GHz

< -15 dBc (at 5 dBm, typical)

3 GHz to 8.5 GHz

< -10 dBc (at 5 dBm, typical)

Non-harmonic spurious

10 MHz to 3 GHz

< -25 dBc (at 5 dBm, typical)

3 GHz to 8.5 GHz

< -10 dBc (at 5 dBm, typical)