Measurement throughput summary – Atec Agilent-E5071B User Manual
Page 19
![background image](/manuals/318376/19/background.png)
19
Table 1-20
Typical cycle time for measurement completion
1, 2
(ms)
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected 4
5
7
18
2-port cal
5
8
13
42
Start 300 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected 11
12
13
23
2-port cal
20
23
25
46
Start 300 kHz, stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected 19
24
24
24
2-port cal
37
46
48
50
Table 1-21
Typical cycle time for measurement completion
1, 3
(ms)
Number of points
51
201
401
1601
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected 4
6
8
22
2-port cal
5
10
16
56
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected 11
12
13
23
2-port cal
20
24
25
55
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected 20
24
24
26
2-port cal
37
46
47
57
Table 1-22
Typical cycle time for measurement completion
1, 4
(ms)
Number of points
51
201
401
1601
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected 7
17
29
90
2-port cal
12
32
55
178
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected 14
27
43
130
2-port cal
26
50
84
258
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected 16
30
49
146
2-port cal
30
57
96
291
Measurement throughput summary
1
Typical performance.
2
Fast swept mode. System error correction OFF. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1.
3
Fast swept mode. System error correction ON. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1.
4
Standard stepped mode. System error correction ON. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1.