Measurement throughput summary – Atec Agilent-E5071B User Manual

Page 19

background image

19

Table 1-20

Typical cycle time for measurement completion

1, 2

(ms)

Number of points

51

201

401

1601

Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth

Uncorrected 4

5

7

18

2-port cal

5

8

13

42

Start 300 kHz, stop 3 GHz, 100 kHz IF bandwidth

Uncorrected 11

12

13

23

2-port cal

20

23

25

46

Start 300 kHz, stop 8.5 GHz, 100 kHz IF bandwidth

Uncorrected 19

24

24

24

2-port cal

37

46

48

50

Table 1-21

Typical cycle time for measurement completion

1, 3

(ms)

Number of points

51

201

401

1601

Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth

Uncorrected 4

6

8

22

2-port cal

5

10

16

56

Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth

Uncorrected 11

12

13

23

2-port cal

20

24

25

55

Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth

Uncorrected 20

24

24

26

2-port cal

37

46

47

57

Table 1-22

Typical cycle time for measurement completion

1, 4

(ms)

Number of points

51

201

401

1601

Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth

Uncorrected 7

17

29

90

2-port cal

12

32

55

178

Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth

Uncorrected 14

27

43

130

2-port cal

26

50

84

258

Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth

Uncorrected 16

30

49

146

2-port cal

30

57

96

291

Measurement throughput summary

1

Typical performance.

2

Fast swept mode. System error correction OFF. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1.

3

Fast swept mode. System error correction ON. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1.

4

Standard stepped mode. System error correction ON. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1.