Table 1-3 Corrected system performance with type-N device connectors, 85092C electronic calibration module

Network analyzer: E5070B/E5071B, calibration module: 85092C (type-N, 50 Ω) electronic calibration (ECal) module, calibration: full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with < 1°C deviation
from calibration temperature, isolation calibration not omitted

Description

Specification

(dB)

3 MHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 8.5 GHz

Directivity 52 52 47

Source match

45

41

36

Load match

47

44

39

Reflection tracking

±0.040

±0.060

±0.070

Transmission tracking

±0.039

±0.069

±0.136

Transmission uncertainty (specification)

Magnitude Phase

Reflection uncertainty (specification)
Magnitude Phase