Table 4-1

Corrected system performance with type-N 75 Ω device connectors, 85036E calibration kit

Network analyzer: E5070B/E5071B, calibration kit: 85036E (type-N 75 Ω), 50 Ω to 75 Ω adapters: 11852B, calibration: full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23° C ± 5° C with < 1° C deviation from
calibration temperature, Isolation calibration not omitted

Description

Supplemental information (dB, typical)

3 MHz to 3 GHz

Directivity 37

Source match

33

Load match

37

Reflection tracking

±0.017

Transmission tracking

±0.021

Transmission uncertainty 3 MHz to 3 GHz (supplemental information, typical)

Magnitude Phase

Reflection uncertainty 3 MHz to 3 GHz (supplemental information, typical)
Magnitude Phase

Corrected system performance for 75 Ω measurements with 11852B
50 Ω to 75 Ω minimum-loss pads (supplemental information)