Table 1-23
Cycle time
1,2
(ms) vs. number of points
1
Number of points
Fast swept mode
Fast swept mode
Standard stepped mode
system error correction OFF
system error correction ON
system error correction ON
3
4 4 4
11
4 4 4
51
4 4 7
101
4
5
11
201
5
6
17
401
8
8
29
801
11 13 52
1601
18 23 90
Table 1-24
Data transfer time
1
(ms)
Number of points
51
201
401
1601
SCPI over GPIB
3
64-bit floating point
5
16
29
109
ASCII 21
79
156
617
SCPI over 100 Mbps LAN (telnet)
3
REAL 64
2
2
3
5
ASCII 34
128
254
995
SCPI over 100 Mbps LAN (SICL-LAN)
3
REAL 64
4
4
5
8
ASCII 6
14
26
95
SCPI over USB (USBTMC)
4
REAL 64
4
5
5
7
ASCII 6
18
33
126
COM (program executed in the analyzer)
5
Variant type
1
1
1
1
1
Typical performance.
2
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth, Error correction OFF, display update: OFF, number of traces = 1.
3
Measured using a VEE 6.0 program running on a 733 MHz Pentium III HP Kayak, Transferred complex S
11
data, using :CALC:DATA?SDATA.
4
Measured using a VEE 7.0 program running on a 500 MHz Pentium III DELL OptiPlex, transferred complex S
11
data.
5
Measured using an E5070B/E5071B VBA macro running inside the analyzer. Transferred complex S
11
data.