Test port input – Atec Agilent-E5071B User Manual
Page 11
11
Table 1-10
Test port input levels
Description Specification
Supplemental
information
Maximum test port input level
300 kHz to 3 GHz
+10 dBm
3 GHz to 4.25 GHz
+9 dBm
3 GHz to 6 GHz
+7 dBm
6 GHz to 8.5 GHz
+5 dBm
Damage level
300 kHz to 8.5 GHz
RF +20 dBm
±10 VDC (source attenuator = 0 dB)
±25 VDC (source attenuator = 5 dB or more, typical)
Crosstalk
1
3 MHz to 3 GHz
-120 dB
3 GHz to 6 GHz
-109 dB
6 GHz to 7.5 GHz
-99 dB
7.5 GHz to 8.5 GHz
-89 dB
Table 1-11
Test port input (trace noise
2
)
Description Specification
Supplemental
information
Trace noise
magnitude
300 kHz to 3 MHz
5 mdB rms (typical)
(source power level = +10 dBm)
8 mdB rms (high temperature mode: ON, typical)
3 MHz to 3 GHz
1 mdB rms (23°C ±5°C)
4 mdB rms
(source power level = +10 dBm)
(high temperature mode: ON, typical)
3 GHz to 4.25 GHz
1.2 mdB rms (23°C ±5°C)
4.8 mdB rms
(source power level = +9 dBm)
(high temperature mode: ON, typical)
4.25 GHz to 6 GHz
3.6 mdB rms (23°C ±5°C)
7.2 mdB rms
(source power level = +7 dBm)
(high temperature mode: ON, typical)
6 GHz to 7.5 GHz
3.6 mdB rms (23°C ±5°C)
7.2 mdB rms
(source power level = +5 dBm)
(high temperature mode: ON, typical)
7.5 GHz to 8.5 GHz
6 mdB rms (23°C ±5°C)
9.6 mdB rms
(source power level = +5 dBm)
(high temperature mode: ON, typical)
Trace noise
phase
300 kHz to 3 MHz
0.035° rms (23°C ±5°C, typical)
(source power level = +10 dBm)
0.05° rms (high temperature mode: ON, typical)
3 MHz to 3 GHz
0.007° rms (23°C ±5°C, typical)
(source power level = +10 dBm)
0.02° rms (high temperature mode: ON, typical)
3 GHz to 4.25 GHz
0.008° rms (23°C ±5°C, typical)
(source power level = +9 dBm)
0.024° rms (high temperature mode: ON, typical)
4.25 GHz to 6 GHz
0.025° rms (23°C ±5°C, typical)
(source power level = +7 dBm)
0.042° rms (high temperature mode: ON, typical)
6 GHz to 7.5 GHz
0.025° rms (23°C ±5°C, typical)
(source power level = +5 dBm)
0.042° rms (high temperature mode: ON, typical)
7.5 GHz to 8.5 GHz
0.042° rms (23°C ±5°C, typical)
(source power level = +5 dBm)
0.06° rms (high temperature mode: ON, typical)
Test port input
1
Response calibration not omitted.
2
Trace noise is defined as a ratio measurement of a through, at IF bandwidth = 3 kHz.