Table 1-4 Corrected system performance with 3.5 mm device connector type, 85033E calibration kit

Network analyzer: E5070B/E5071B, calibration kit: 85033E (3.5 mm, 50 Ω), calibration: full 2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23°C ± 5°C with < 1°C deviation from
calibration temperature, isolation calibration not omitted

Description

Specification

(dB)

3 MHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 8.5 GHz

Directivity 46 38 38

Source

match

43 37 36

Load

match

46 38 38

Reflection

tracking

±0.006 ±0.009 ±0.010

Transmission

tracking

±0.016 ±0.065 ±0.079

Transmission uncertainty (specification)

Magnitude Phase

Reflection uncertainty (specification)