beautypg.com

E5091a multiport test set – Atec Agilent-E5071B User Manual

Page 26

background image

26

Table 2-2

Option E5091A-009 port performance

Description

Specification

50 MHz to 300 MHz

300 MHz to 1.3 GHz

1.3 GHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 8.5 GHz

Load match

Test port selected

A, T2, R1+, R1–

19 dB

20 dB

18 dB

12 dB

10 dB

T1, R2+, R2–, R3+, R3–

15 dB

17 dB

15 dB

11 dB

8 dB

Test port unselected

A, T2, R1+, R1–, R3+, R3–

23 dB

25 dB

19 dB

12 dB

11 dB

T1, R2+, R2–

18 dB

20 dB

16 dB

12 dB

9 dB

Interconnect port, typical

P1, P2, P3, P4

19 dB

19 dB

17 dB

13 dB

9 dB

Insertion loss

Test port

A, T2, R1+, R1–

3 dB

3 dB

4 dB

5 dB

6 dB

T1, R2+, R2–, R3+, R3–

5 dB

5 dB

7 dB

8 dB

9.5 dB

Stability, typical

0.005 dB/° C

0.005 dB/° C

0.005 dB/° C

0.01 dB/° C

0.015 dB/° C

Isolation

Over arbitrarily test ports

-100 dB

-100 dB

-100 dB

-100 dB

-90 dB

E5091A multiport test set

The section provides test set input/output performance without calibration by the E5070B/E5071B.

Table 2-1

Test set input/output performance

Description Specification Supplemental

information

Range

50 MHz to 8.5 GHz

Damage level

20 dBm, ±25 VDC (typical)