Corrected system performance – Atec Agilent-E5071B User Manual
Page 3
![background image](/manuals/318376/3/background.png)
3
Corrected system performance
The specifications in this section apply for measurements
made with the Agilent E5070B/E5071B network analyzer
with the following conditions:
• No averaging applied to data
• Environmental temperature of 23°C ± 5°C, with less
than 1°C deviation from the calibration temperature
• Response and isolation calibration not omitted
Table 1-1
System dynamic range
1,2
Description Specification
Supplemental
information
System dynamic range
300 kHz to 3 MHz, IF bandwidth = 3 kHz
85 dB
3 MHz to 1.5 GHz, IF bandwidth = 3 kHz
95 dB
98 dB
1.5 GHz to 3 GHz, IF bandwidth = 3 kHz
97 dB
100 dB
3 GHz to 4 GHz, IF bandwidth = 3 kHz
96 dB
99 dB
4 GHz to 6 GHz, IF bandwidth = 3 kHz
92 dB
94 dB
6 GHz to 7.5 GHz, IF bandwidth = 3 kHz
87 dB
90 dB
7.5 GHz to 8.5 GHz, IF bandwidth = 3 kHz
80 dB
83 dB
300 kHz to 3 MHz, IF bandwidth = 10 Hz
110 dB
3 MHz to 1.5 GHz, IF bandwidth = 10 Hz
120 dB
123 dB
1.5 GHz to 3 GHz, IF bandwidth = 10 Hz
122 dB
125 dB
3 GHz to 4 GHz, IF bandwidth = 10 kHz
121 dB
124 dB
4 GHz to 6 GHz, IF bandwidth = 10 Hz
117 dB
119 dB
6 GHz to 7.5 GHz, IF bandwidth = 10 Hz
112 dB
115 dB
7.5 GHz to 8.5 GHz, IF bandwidth = 10 Hz
105 dB
108 dB
1
The test port dynamic range is calculated as the difference between the test port rms noise floor
and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.
2
May be limited to 90 dB at particular frequencies below 350 MHz or above 4.25 GHz due to
spurious receiver residuals.