Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-88

determining the BGA-PC content of algal suspensions by using standard cell counting techniques. See

Section 5.17, Principles of Operation of this manual for more information about BGA-PC standards.

Select the 2-point option to calibrate the BGA-PE probe using two calibration standards. In this case, one

of the standards must be clear water (0 cells/mL) and the other should be in the range of the suspected

BGA-PC content at the environmental site.

To begin the calibration, immerse the sonde in the 0 µg/L standard, as instructed, and press Enter. It is

mandatory that the 0 cells/mL standard be calibrated first. The screen will display real-time readings that

will allow you to determine when the readings have stabilized. Pressing Enter will confirm the first

calibration. Following the instructions on the screen, place the sonde in the second BGA-PC standard,

input the correct value in cells/mL, press Enter, and view the stabilization of the values on the screen in

real-time. After the readings have stabilized, press Enter to confirm the calibration. Then, as instructed,

press any key to return to the Calibrate menu.

For all BGA-PE calibration procedures, be certain that the standard and sensor are thermally equilibrated

prior to proceeding with the calibration

CAUTION: To be assured of the accuracy for field measurements made with your BGA-PE sensor, you

must either initially calibrate the sensor using an algal suspension of known BGA-PE content or post

calibrate your sensor in a spreadsheet by comparing the fluorescence values obtained in field studies with

those obtained by laboratory analysis for chlorophyll on grab samples collected during the field study. The

use of chemical dye standards for “calibration” of the sensor may not result in accurate correlation between

fluorescence and actual BGA-PE readings and is designed primarily to evaluate the sensor for drift. See

Section 5.17. Principles of Operation for more information on the proper calibration of your chlorophyll

sensor.



OPTIC RHODAMINE WT

When selecting Rhodamine, there will be a choice of 1-point, 2-point, or 3-point calibrations for your

rhodamine WT sensor.

The 1-point option is normally used to zero the rhodamine probe in 0 ug/L standard. Place the sonde in

clear water with no suspended solids, and input 0 ug/L at the screen prompt. Press Enter and the screen

will display real-time readings that will allow you to determine when the rhodamine readings have

stabilized. Press Enter after the readings have stabilized to confirm the calibration and zero the sensor.

Then, as instructed, press any key to return to the Calibrate menu.

Select the 2-point option to calibrate the rhodamine probe using only two calibration standards. In this

case, one of the standards must be clear water (0 ug/L) and the other should be in the range of the predicted

rhodamine concentration projected for the study. For general purpose measurements an appropriate choice

of standards is usually 0 and 100 ug/L.

To begin the calibration, immerse the sonde in the 0 ug/L standard, as instructed, and press Enter. It is

mandatory that the 0ug/L standard be calibrated first. The screen will display real-time readings that will

allow you to determine when the readings have stabilized. Pressing Enter will confirm the first calibration.

Following the instructions on the screen, place the sonde in the second rhodamine standard, input the

correct rhodamine value in ug/L, press Enter, and view the stabilization of the values on the screen in real-

time. After the readings have stabilized, press Enter to confirm the calibration. Then, as instructed, press

any key to return to the Calibrate menu.

Select the 3-point option for maximum accuracy over the entire range of 0 to 200 ug/L. As for the 2-point

procedure, one of the standards must be 0 NTU. The procedure for this calibration is the same as for a 2-