650 MDS

Section 3

YSI Environmental Monitoring Systems Operations Manual

3-16

“click”. After connecting the sonde to the 650, press the “on” button on the upper left of the 650 keypad to

activate the 650 Main menu as shown below:

Using the arrow keys, highlight the Sonde run entry in the Main 650 Menu and press the Enter key.

Readings from the 6-series sonde sensors will appear on the 650 Main Display in real-time as shown in the

following diagram.

The two windows at the top of the display give the capability of logging the displayed readings to either the

650 (top left) or the sonde (top right) memory and allow the user to activate the wiper on turbidity,

chlorophyll, and rhodamine WT probes if they are present in the sonde. Instructions for use of these

windows are provided in subsequent sections of this manual. Note also that there is a disk with rotating

segments in the upper right corner of the display whose movement confirms that the 650/sonde interface is

functional.

To return to the 650 Main menu, press the Escape key.

CAUTION: When using the 650 MDS in field applications which are in close proximity to communication

towers and heavy industrial equipment, or which involve very low humidity conditions, the user may

experience problems with logger function. If possible, the user should relocate the sampling site away

from these potential sources of interference. In most cases, symptoms will involve instrument shutdown or

display lockup. These problems can usually be overcome by either simply pressing the on/off key to

reactivate the meter display or by resetting the instrument by removal of battery power as described in