Specifications

Appendix O

YSI Incorporated

Environmental Monitoring Systems Operations Manual

O-10

Non-Vented Level - Shallow

Sensor Type..............Stainless steel strain gauge

Range........................0 to 30 ft (9.1 m)

Accuracy ................ +/- 0.06 ft (0.018 m)

Resolution.................0.001 ft (0.001 m)

Temperature Range -5 to 45 C

Vented Level - Shallow

Sensor Type..............Stainless steel strain gauge

Range........................0 to 30 ft (9.1 m)

Accuracy, 0-30ft ......+/- 0.01 ft (0.003 m)

Resolution.................0.001 ft (0.001 m)

Temperature Range -5 to 45 C

Temperature

Sensor Type...........Thermistor

Range.....................-5 to 50

C

Accuracy................+/- 0.15

C

Resolution..............0.01

C

Depth……………200 meters

Rapid Pulse Dissolved Oxygen, % saturation

Sensor Type...........Rapid Pulse - Clark type, polarographic

Range.....................0 to 500 % air saturation

Accuracy................0-200 % air saturation, +/- 2 % of the reading or 2 % air saturation, whichever is greater

200-500 % air saturation, +/- 6 % of reading

Resolution..............0.1 % air saturation

Temperature Range -5 to 50 C

Depth……………200 meters

Rapid Pulse Dissolved Oxygen, mg/L (Calculated from % air saturation, temperature and salinity)

Sensor Type.......... Rapid Pulse - Clark type, polarographic

Range.....................0 to 50 mg/L

Accuracy................0 to 20 mg/L, +/- 2 % of the reading or 0.2 mg/L, whichever is greater

20 to 50 mg/L, +/- 6 % of the reading

Resolution..............0.01 mg/L

Temperature Range -5 to 50 C

Depth……………200 meters

ROX Optical Dissolved Oxygen, % saturation

Sensor Type...........Optical, Luminescence Lifetime

Range.....................0 to 500 % air saturation

Accuracy ...............0-200 % air saturation, +/- 1 % of the reading or 1 % air saturation, whichever is greater

200-500 % air saturation, +/- 15 % of reading; Relative to Calibration Gases.

Resolution..............0.1 % air saturation

Temperature Range -5 to 50 C

Depth…………… 61 meters

ROX Optical Dissolved Oxygen, mg/L (Calculated from % air saturation, temperature and salinity)

Sensor Type.......... Optical, Luminescence Lifetime

Range.....................0 to 50 mg/L

Accuracy................0 to 20 mg/L, +/- 1 % of the reading or 0.1 mg/L, whichever is greater

20 to 50 mg/L, +/- 15 % of the reading; Relative to Calibration Gases.

Resolution..............0.01 mg/L

Temperature Range -5 to 50 C

Depth…………… 61 meters