beautypg.com

Xylem 6-Series Multiparameter User Manual

Page 51

background image

Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-45

CALIBRATION TIP: Exposure to the high ionic content of pH buffers can cause a significant, but

temporary, drift in these ISE probes (ammonium, nitrate and chloride probes). Therefore, when calibrating

the pH probe, YSI recommends that you use one of the following methods to minimize errors in the

subsequent readings:

Calibrate pH first, immersing all of the probes in the pH buffers. After calibrating pH, place the

probes in 100 mg/L nitrate or ammonium standard or 1000 mg/L chloride standard and monitor the

reading. Usually, the reading starts low and may take as long as 30 minutes to reach a stable value.

When it does, proceed with the calibration.

When calibrating pH, remove ISE modules from the sonde bulkhead and plug the ports. After pH

calibration is complete, replace the ISE sensors and proceed with their calibration with no stabilization

delay.

TURBIDITY 2-POINT

Select Optic X –Turbidity-6136 or Optic X – Turbidity 6026 from the Calibrate Menu and then 2-2-

Point.

NOTE: Before calibrating your 6136 turbidity sensor, pay particular attention to the following cautions:

To properly calibrate YSI turbidity sensors, you MUST use standards that have been prepared

according to details in Standard Methods for the Treatment of Water and Wastewater (Section 2130 B).

Acceptable standards include (a) formazin prepared according to Standard Methods; (b) dilutions of

4000 NTU formazin concentrate purchased from Hach; (c) Hach StablCal

TM

standards in various NTU

denominations; and (d) AMCO-AEPA standards prepared specifically for the 6026 and 6136 by either

YSI or approved vendors who are listed on the YSI website (www.ysi.com). STANDARDS FROM

OTHER VENDORS ARE NOT APPROVED FOR THE YSI TURBIDITY SYSTEM AND THEIR

USE WILL LIKELY RESULT IN BOTH CALIBRATION ERRORS AND INCORRECT FIELD

READINGS.

For AMCO-AEPA standards, the value entered by the user during the calibration protocol is

DIFFERENT depending on which YSI turbidity sensor (Legacy 6026 or Current 6136) is being

calibrated. This reflects the empirically determined fact that 6026 and 6136 sensors that have been

calibrated to the same value in the primary standard formazin, will have different responses in

suspensions of the AEPA-AMCO beads. This effect is likely due to the larger optical cell volume of

the 6136. Thus, for example, the label of the YSI 6073 turbidity standard bottle indicates that the

value of the standard is 100 NTU when used for calibration of the 6026 sensor, but 126 NTU when

used to calibrate the 6136. Note that the phenomenon of a sensor-specific formazin/AEPA-AMCO

ratio is well known for sensors other than the 6026 and 6136.

When calibrating a 6136 turbidity sensor you MUST carefully follow the instructions found above in

Section 2.6.1 to avoid interference from the bottom of the calibration cup. Failure to carry out the

calibration properly can result in inaccurate readings, particularly water of very low turbidity.

Before proceeding with the calibration, be certain that the probe compartment of the sonde has been

cleaned and is free of debris. Solid particles from this source, particularly those carried over from past

deployments, will contaminate the standards during your calibration protocol and cause either

calibration errors and/or inaccurate field data.

One standard must be 0 NTU, and this standard must be calibrated first.

To begin the calibration, place the correct amount (see Tables 1-8 above) of 0 NTU standard (clear

deionized or distilled water) into the calibration cup provided with your sonde. Immerse the sonde in the

This manual is related to the following products: