When selecting ISE3-NH4+, you will be given the choice of 1-point, 2-point, or 3-point calibrations for

your ammonium (NH

4

+

) sensor.

Select the 1-point option only if you are adjusting a previous calibration. If a 2-point or 3-point calibration

has been performed previously, you can adjust the calibration by doing a one point calibration. Immerse

the sonde in any solution of known ammonium concentration and press Enter. You will be prompted to

type in the NH

4

+

value (in mg/L of NH

4

-N) of the solution you are using. Press Enter again, and the screen

will display real-time readings that will allow you to determine when the NH4

+

readings have stabilized.

Pressing Enter will confirm the calibration.

Select the 2-point option to calibrate the NH

4

+

probe using only two calibration standards that are both at

approximately the temperature of your environmental sample. In this procedure, the NH

4

+

sensor is usually

calibrated using solutions that contain 1 and 100 mg/L of NH

4

-N . Be certain that the calibration solution

and sensor are thermally equilibrated prior to entering NH

4

+

values.

To begin the calibration immerse the sonde in the 1 mg/L standard, press Enter, input the NH

4

-N value,

and again press Enter. The screen will display real-time readings that will allow you to determine when

the sensor has stabilized. Pressing Enter will confirm the first calibration. Following the instructions on

the screen, place the sonde in the second NH

4

+

standard, press Enter, input the correct concentration value,

again press Enter, and view the stabilization of the values on the screen in real time. After the readings

have stabilized, press Enter to confirm the calibration. Then, as instructed, press any key to return to the

Calibrate menu.

Select the 3-point option to calibrate the NH

4

+

probe using three calibration solutions, two at ambient

temperature and one at a temperature significantly different from ambient. The 3-point calibration method

should be used to assure maximum accuracy when the temperature of the media to be monitored cannot be

anticipated. The procedure for this calibration is the same as for a 2-point calibration, but the software will

prompt you to place the sonde in the additional solution to complete the 3-point procedure. Be certain that

the calibration solution and sensor are thermally equilibrated prior to proceeding with the calibration. The

recommended order of calibration standards is (1) 100 mg/L standard at ambient temperature, (2) 1 mg/L

standard at ambient temperature, and (3) 1 mg/L standard at a different temperature (usually lower) than

ambient. For best results, insure a temperature difference of at least 10 C

between the two 1 mg/L

standards.

NOTE: YSI strongly recommends the use of the 3-point protocol to ensure the best possible performance

from all ISE sensors (ammonium, nitrate, and chloride)


NITRATE

When selecting ISE4 NO3, you will be given the choice of 1-point, 2-point, or 3-point calibrations for your

nitrate (NO

3

-

) sensor. The procedure is identical to that for the ammonium sensor, except that the calibrant

values are in mg/L of NO

3

-N instead of NH

4

-N.

NOTE: YSI strongly recommends the use of the 3-point protocol to ensure the best possible performance

from all ISE sensors (ammonium, nitrate, and chloride)