beautypg.com

Xylem 6-Series Multiparameter User Manual

Page 150

background image

650 MDS

Section 3

YSI Environmental Monitoring Systems Operations Manual

3-20

3.3.7 ID

Activate this item and Enter choice of instrument identification (limit 11 characters) for your 650 as

described in Section 3.1.8 above.

3.3.8 Circuit board SN

This item identifies the serial number of the circuit board resident in your 650. It cannot be changed by the

user.

3.3.9 Barometer Units:

Highlighting this selection and pressing Enter will activate a display which will allow you to select the

units of barometric pressure which will appear in real-time in the display Status Bar. Highlight the desired

units and press Enter to confirm the selection. This entry will only appear if you have ordered your 650

with the barometer option.

3.3.10 Calibrate Barometer

You may need to scroll down with the arrow keys to view this selection which allows you to calibrate the

optional barometer to your local barometric pressure. While the barometer has been factory calibrated and

should provide accurate readings on receipt, some sensor drift may occur over time, requiring occasional

calibration by the user. Proceed as follows:

Determine your local barometric pressure. This value is usually obtained from an independent laboratory

barometer. You can also obtain a barometric pressure value from your local weather service. This latter

value has usually been corrected to sea level and therefore must be “uncorrected” before it can be used in

calibration of the 650 barometer. See Section 3.9 of this manual for details.

Select Calibrate barometer and press Enter. Input the known barometric pressure from the 650 keypad

in the selected units (mm Hg in the following example) and press Enter. Note that in the example, the

barometer output has changed from 733.4 (the previous incorrect value) to 740 (the correct value from your

laboratory barometer). Note also that a value of the approximate barometer offset which occurred during

the calibration (6.3 mm Hg) is present in the display. The offset value is approximate due to the variation

in the tenths of a mm digit of the sensor. This value can be used to return the sensor to its factory

calibration; simply recalibrate the sensor to a new value which is determined by (current value) – (offset).

Note that this entry will only appear if your have ordered your 650 with the barometer option.

NOTE: After initial calibration of the barometer, a file with a name defined by the Circuit Board ID and a

.glp extension (e.g., 00003245.glp) will automatically be generated and stored in 650 memory. Each

additional barometer calibration record for will also be stored. See Section 6.3 below for the transfer of

this file to a computer for viewing and storage.

This manual is related to the following products: