2.3 PREPARING THE SONDE FOR USE

To prepare the sonde for calibration and operation, you need to install probes (sensors) into the connectors

on the sonde bulkhead. In addition to probe installation, you need to install a new membrane on the YSI

6562 DO Probe if you are using this item. It is recommended that you install the DO membrane before

installing the probe onto the bulkhead. For membrane changes in the future, you may be able to perform

this operation without removing the DO probe. This will largely depend on whether the other installed

probes interfere with your ability to install a membrane. The next step is providing power for the sondes,

through batteries or line power, and then connecting a field cable. The four steps necessary for getting your

sonde ready for use are listed below.

Step 1 Installing the Dissolved Oxygen Membrane – Section 2.3.1

Step 2 Installing the Probes – Section 2.3.2

Step 3 Supplying Power – Section 2.3.3

Step 4 Connecting a Field Cable – Section 2.3.4

2.3.1 STEP 1 - INSTALLING THE DISSOLVED OXYGEN MEMBRANE

Note: If you are using a ROX Optical DO sensor for your applications, please skip to Section 2.3.2 at

this time.

The 6562 Rapid Pulse Polarographic DO probe is shipped with a protective dry membrane on the sensor tip

held in place by an O-ring. Remove the O-ring and membrane. Handle the probe with care. It is very

important not to scratch or contaminate the sensor tip. See Section 2.10.2, Probe Care and Maintenance,

for information on how often the membrane should be replaced.


Unpack the YSI 6562 DO Probe Kit and follow the instructions below

.

Open the membrane kit and prepare the electrolyte solution. Dissolve the KCl in the dropper bottle by

filling it to the neck with deionized or distilled water and shaking until the solids are fully dissolved. After

the KCl is dissolved, wait a few minutes until the solution is free of bubbles.

Figure 5 Figure 6


The DO membrane can be installed with the DO probe either free or installed in the sonde. Both methods

are described in detail below. CAUTION: If you install the membrane with the probe not installed in

the sonde, be sure that the protective cap is installed on the probe end away from the sensor face to

ensure that the connector is not contaminated with electrolyte.

1.

ADD DI OR DISTILLED

WATER

2.

PROTECTIVE CAP

DRY MEMBRANE