beautypg.com

Effect of turbidity on bga-pc readings – Xylem 6-Series Multiparameter User Manual

Page 272

background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-34

2. Accurately weigh 0.500 g of the 20 % Rhodamine concentrate, quantitatively transfer the viscous liquid

to a 1000-mL volumetric flask and fill the flask to the top graduation. Mix well. This solution contains

100 mg of Rhodamine WT per 1000 mL of water.

3. Accurately transfer 1.0 mL of the solution prepared in the above step to a 1000 mL volumetric and then

fill the flask to the top graduation with purified water. Mix well to obtain a solution that is 100 ug/L

(0.10 mg/L) in water (a 1000:1 dilution of the concentrated solution).

4. Store the concentrated standard solution in a darkened glass bottle in a refrigerator to retard

decomposition. The dilute standard prepared in the previous step should be used within 5 days of its

preparation.

When Rhodamine WT standards are required in the future, perform another dilution of the concentrated dye

solution after warming it to ambient temperature. Our experience has indicated that the concentrated solution

that has been kept at cold temperatures is much more stable than the dilute solution stored at room

temperature.

It is well known that the intensity of the fluorescence of many dyes shows an inverse relationship with

temperature. The effect must also be accounted for when “calibrating” the YSI PC sensor with Rhodamine

WT. Enter the calibration value from the table below corresponding to the temperature of the standard.

WARNING: The “PC tempco” factor in the Advanced|Sensor menu MUST BE SET TO ZERO,

when calibrating with Rhodamine WT.

Table. Approximate PC-containing equivalent of 100 ug/L Rhodamine WT as a function of temperature.

T, C Cells/mL to Enter

T,

C

Cells/mL to Enter

30

44940

18

76580

28

49700

16

83580

26

54600

14

91420

24

58940

12

98140

22

64120

10

107940

20

70000

8

113540

REMEMBER: The use of Rhodamine WT for “calibration” of the Phycocyanin Probe is only an

approximation. To assure the maximum accuracy for the 6131 sensor, the user must relate the field

fluorescence readings to data from actual BGA samples as described above. YSI does not provide an

accuracy specification for the sensor due to the limitations described above and below.

EFFECT OF TURBIDITY ON BGA-PC READINGS

As described above, the filters in front of the photodiode in the YSI 6131 Phycocyanin Probe prevent most

of the orange light which is used to excite the Rhodamine molecules from reaching the detector after being

backscattered off of non-fluorescent particles (turbidity) in environmental water. However, the filter

system is not perfect and a minor interference on PC-containing BGA readings from suspended solids will

result. Laboratory experiments indicate that a YSI 6131 sensor will have a turbidity interference

characterized by a factor of about 21 cells/mL of PC-containing BGA per NTU of turbidity. For example,

at a turbidity of 100 NTU, a PC-containing BGA reading of 2100 cell/mL will be observed over and above

the reading actually due to the presence of BGA. Users may wish to use the independently-determined

turbidity value and the above compensation factor to correct measured PC-containing BGA values using,

for example, a spreadsheet.

This manual is related to the following products: