Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-18

sonde software to provide readings in nephelometric turbidity units (NTUs). A schematic of a YSI

turbidity sensor is shown in the following picture.

Two turbidity probes have been available for use with YSI 6-series sondes – the 6026 and the 6136. The

6026 probe is no longer available from YSI, but since a large number of this probe type are still in use, a

description of the sensor will be included in this section. The probes are detailed briefly below; both are

equipped with a mechanical wiper to periodically clean the sensor either by manual or automatic activation.

These wiper systems make the probe ideal for long term monitoring, but they also work well for spot

sampling applications.

The Model 6026 probe was offered by YSI from 1995 to approximately 2002 and is characterized by

relatively small optics, a factor that results in minimal penetration of the light beam into the sample

and thus allows the use of shorter probe guards on the sonde in which it is installed. If a 6026 sensor

is selected in the Sensor menu of the sonde, the turbidity will automatically be reported in the units

“turbid NTU”.

The Model 6136 has been the only turbidity probe sold by YSI since 2002 and it is characterized by

relatively large optics compared to the 6026. Since these larger optics result in deeper penetration of

the light beam into the sample, the use of a longer probe guard is required, a slight disadvantage.

However, the larger optical cell volume of the sensor has two significant advantages: (1) There is less

background noise associated with the turbidity readings and (2) the absolute values of the readings are

significantly closer to those from typical laboratory turbidimeters (such as the Hach 2100AN) which

use a large cell volume. If a 6136 sensor is selected in the Sensor menu of the sonde, the turbidity will

automatically be reported in the units “turbid+ NTU”.

In the primary standard formazin, the two turbidity sensors (6026 and 6136) will show effectively identical

behavior. In addition, AEPA-AMCO polymer beads (supplied by YSI as Model 6073) can also be used

with one important qualification. You will need to remember that the 6026 and 6136 sensors, which read

the same in formazin suspensions, will have different responses in the suspensions of the AEPA-AMCO

beads. This effect is due to the larger optical cell volume of the 6136. Thus, as noted on the label of the

Optical Fiber

Photodetector

Light Source

Near Infrared LED