beautypg.com

Effect of turbidity on bga-pe readings – Xylem 6-Series Multiparameter User Manual

Page 278

background image

Principles of Operation

Section 5

YSI Incorporated

Environmental Monitoring Systems Manual

5-40

2. Accurately weigh 0.500 g of the 20 % Rhodamine concentrate, quantitatively transfer the viscous liquid

to a 1000-mL volumetric flask and fill the flask to the top graduation. Mix well. This solution contains

100 mg of Rhodamine WT per 1000 mL of water.

3. Accurately transfer 80 uL of the solution prepared in the above step to a 1000 mL volumetric and then fill

the flask to the top graduation with purified water. Mix well to obtain a solution that is 8 ug/L (0.008

mg/L) in water.

4. Store the concentrated standard solution in a darkened glass bottle in a refrigerator to retard

decomposition. The dilute standard prepared in the previous step should be used within 5 days of its

preparation.

When Rhodamine WT standards are required in the future, perform another dilution of the concentrated dye

solution after warming it to ambient temperature. Our experience has indicated that the concentrated solution

that has been kept at cold temperatures is much more stable than the dilute solution stored at room

temperature.

It is well known that the intensity of the fluorescence of many dyes shows an inverse relationship with

temperature. The effect must also be accounted for when “calibrating” the YSI BGA-PE sensor with

Rhodamine WT. Enter the calibration value from the table below corresponding to the temperature of the

standard.

WARNING: The “PE tempco” factor in the Advanced|Sensor menu MUST BE SET TO ZERO,

when calibrating with Rhodamine WT.

Table. Approximate PE-containing equivalent of 8 ug/L Rhodamine WT as a function of temperature.


T, C Cells/mL to Enter

T,

C

Cells/mL to Enter

30

156,000

18

210,000

28

164,000

16

220,000

26

174,000

14

230,000

24

181,000

12

240,000

22

189,000

10

247,000

20

200,000

8

254,000

REMEMBER: The use of Rhodamine WT for “calibration” of the PE Probe is only an approximation.

To assure the maximum accuracy for the 6132 sensor, the user must relate the field fluorescence

readings to data from actual BGA samples as described above. YSI does not provide an accuracy

specification for the sensor due to the limitations described above and below.

EFFECT OF TURBIDITY ON BGA-PE READINGS

As described above, the filters in front of the photodiode in the YSI 6132 PE Probe prevent most of the

orange light which is used to excite the Rhodamine molecules from reaching the detector after being

backscattered off of non-fluorescent particles (turbidity) in environmental water. However, the filter

system is not perfect and a minor interference on PE-containing BGA readings from suspended solids will

result. Laboratory experiments indicate that a YSI 6132 sensor will have a turbidity interference

characterized by a factor of about 140 cells/mL of PC-containing BGA per NTU of turbidity. For example,

at a turbidity of 100 NTU, a PE-containing BGA reading of 14000 cell/mL will be observed over and above

the reading actually due to the presence of BGA. Users may wish to use the independently-determined

turbidity value and the above compensation factor to correct measured PE-containing BGA values using,

for example, a spreadsheet.

This manual is related to the following products: