Software specifications – Xylem 6-Series Multiparameter User Manual
Page 376

Specifications
Appendix O
YSI Incorporated
Environmental Monitoring Systems Operations Manual
O-13
Ammonia-Nitrogen
Sensor Type...........Calculated from ammonium, pH and temperature
Range.....................0-200 mg/L-N
Accuracy................+/- 10% of reading or 2 mg/L (whichever is greater)
Resolution..............0.001 mg/L-N to 1 mg/L-N (range dependent)
Temperature Range -5 to 50 C
Depth……………15 meters
Chloride
Sensor Type...........Solid state ion-selective electrode
Range.....................0-1000 mg/L
Accuracy................+/- 15% of reading or 5 mg/L (whichever is greater)
Resolution..............0.001 mg/L to 1 mg/L (range dependent)
Temperature Range -5 to 50 C
Depth……………15 meters
SOFTWARE SPECIFICATIONS
EcoWatch for Windows (included)
IBM PC compatible computer with CD ROM drive and with a 386 processor (or better) running
Windows software. Compatible with Windows 3.1, Windows 95, Windows 98, Windows ME,
Windows NT, Windows 2000, and Windows XP. Minimum RAM requirement: 4 megabytes