Xylem 6-Series Multiparameter User Manual
Page 109

Sondes
Section 2
YSI Incorporated
Environmental Monitoring Systems Operations Manual
2-103
Turb+ A2
1000
0.6 to 1.5
Range is ratio of (M2-M1) to (A2-A1)
Turb+ M2
Chl Offset
0
-30 to 20
Chl A1
500
0.6 to 1.5
Range is ratio of M1 to A1
Chl M1
500
Chl A2
1000
0.6 to 1.5
Range is ratio of (M2-M1) to (A2-A1)
Chl M2
1000
Chl RFU Offset 0
-0.1 to +0.1
PC Offset
0
-5000 to 10000
PC Gain 1
Not checked
PC RFU Offset 0 Not checked
PE Offset
0
-5000 to 15000
PE Gain 1
Not checked
PE RFU Offset 0
Not checked
Rhod Offset
0
-10 to 10
Rhod A1
500
0.6 to 1.5
Range ratio of M1 to A1
Rhod M1
500
Rhod A2
1000
0.6 to 1.5
Range is ratio of (M2-M1) to (A2-A1)
Rhod M2
1000
To reset a calibration cell constant, access the sonde Calibrate menu. Then select the sensor and type
“UNCAL” instead of the value. This action will change the calibration constants of that sensor back to the
factory default.
From the Advanced menu, select 2-Setup to display miscellaneous options. Type the appropriate number to
activate/deactivate any of the displayed features.
1-(*)VT100 emulation. Activate this option for VT100 terminal emulation. This feature allows the sonde
to send escape sequences to clear the screen which in turn results in an improved display. Usually this
feature should be activated, but, if your terminal or terminal emulator shows odd characters at the
beginning of each menu title, then turn this item off. With the feature off, the sonde will send several
carriage returns and line feeds to 'clear' the display. The number of
page length setting.
2- Start up. This option allows the user to select the mode of sonde operation when power is applied or a
“reset” command is issued. After pressing the “2” key, the following options will be displayed.
------------Advanced setup--------
1-(*)VT100 emulation
2-Start up:Normal
3-( )Comma radix
4-( )Auto sleep RS232
5-( )Auto sleep SDI12
6-( )Multi SDI12
7-( )Full SDI12
8-( )Sample and hold