Sampling Optics: WS-1 Diffuse Reflectance Standard

The WS-1 D

IFFUSE

R

EFLECTANCE

S

TANDARD

is a compact physical standard for use in performing

diffuse reflectance measurements, especially color analysis. The reflectance material that comprises the
WS-1 is Spectralon

, a highly Lambertian thermoplastic substance that provides a nearly 100% diffuse

reflective surface for applications from 200 nm to 2.5

µ

m. (With an Ocean Optics spectrometer, the

practical use of the WS-1 is limited to 200-1100 nm.) The reflective area of the WS-1 is 1.25” in diameter.
The diffusing material is encased in a small, black receptacle with a screw-on top.

Operation

Reflection is the return of radiation by a surface, without change in wavelength. Reflection is expressed as
a percentage relative to the reflection from a standard substance, such as the WS-1 white reference
material. The WS-1 is mostly used in a setup with a reflection probe and reflection probe holder for diffuse
reflection measurements or with the ISP-REF Integrating Sphere. To use the WS-1 as a reference in a
diffuse reflection measurement:
1. Make sure you are in Scope Mode. Take a reference spectrum by first making sure nothing is blocking

the light path going to your reference. Place the reflection probe over the WS-1. For best results, use
our Reflection Probe Holder to keep the reflection probe firmly at a 45

°

angle.

2. Take the reference reading by clicking the Store Reference Spectrum icon on the toolbar or selecting

File | Store Reference Spectrum.

Specifications

Reflectance material:

Spectralon

Reflectivity:

>99%

Wavelength range:

200 nm - 2.5

µ

m*

Reflective area diameter:

1.25”

*