Installing Filters
1. Loosen the filter clamping screw with the provided screwdriver.
2. Insert the filter into the filter slot. The filter slot can accommodate filters up to 6 mm thick.
3. Clamp the filter in place by gently tightening the clamping screw finger-tight.
Using the Temperature Stabilization Feature
This feature is used to heat or cool the cuvette holder base and cuvette.
1. Remove the two plugs from the top side of the base. (The plug on the right side of the base should stay
in place but might require thread tape.)
2. Replace the plugs with the two barbed fittings (or any 1/8” NPT pipe thread fittings). Thread tape
might be required on the fittings to prevent leakage.
3. Connect the fittings to a water source. Water will circulate through the base.
Specifications
Path length:
1 cm
Collimating lenses (VIS-NIR):
(2) BK 7 glass, 5 mm diameter, f/2
Collimating lenses (UV-VIS-NIR):
(2) Dynasil 1100 quartz, 5 mm diameter, f/2
Collimating lens termination:
SMA 905
Filter slot:
accepts filters up to ¼
”
(6 mm) in thickness
Base material:
aluminum
Water input fittings:
1/8
”