Sampling chambers – Ocean Optics S1024DW Install User Manual
Page 25
22
Sampling Chambers
Ocean Optics offers a comprehensive line of compact, low-cost sampling chambers for a variety of UV-
VIS-Shortwave NIR applications. All sampling chambers have SMA terminations for easy coupling to
optical fibers. Options include a variety of cuvette holders for 1-cm and 10-cm cuvettes, in-line filter
holders, flow cells, and other sampling devices.
Standard cuvette holders are available for 1-cm and 10-cm cuvettes,
and couple via SMA-terminated optical fibers to our spectrometers
and light sources to create small-footprint systems for absorbance
experiments. Each cuvette holder comes with a pair of 5-mm diameter
f/2 collimating lenses optimized for UV-VIS-NIR or VIS-NIR
applications, and has a slot for optical filters. Also available is the
CUV-FL-DA, a 1-cm cuvette holder that attaches directly to
light sources.
The CUV-ALL
is a cuvette holder for 1-cm cuvettes that has fiber optic
couplings at four collimators -- a pair of 5-mm diameter f/2 collimating
lenses optimized for the UV-VIS and a pair optimized for the VIS-NIR.
Optical fibers can be attached at these collimators to read or illuminate
the sample. When combined with our spectrometers and light sources,
the CUV-ALL can measure absorbance, fluorescence, scattering, or any
combination of these optical phenomena.
ISS Integrated Sampling System consists of a compact tungsten-halogen light
source with integrated cuvette holder for 1-cm cuvettes. This versatile sampling
device couples to our miniature spectrometers via optical fibers to create fully
integrated, small-footprint sampling systems for VIS-NIR absorbance and
transmission measurements. The ISS is designed specially for use with
S1024DW Spectrometers.
FHS In-Line Filter Holders for the UV-VIS-NIR or VIS-NIR
provide a convenient platform for rapid absorbance or transmission
measurements of optical and other filters. Filter holders come with a
pair of 5-mm diameter collimating lenses, and accept 1” round or
any size square optical filters, up to 6 mm in thickness. FHS
assemblies are typically used either to measure filters or to put
filters into optical setups.
LPC Long Pass Flow Cells couple to our fiber optic spectrometers and
compact light sources to measure low-concentration (ppb-ppt) aqueous
samples. LPC Flow Cells are available in 1-meter, 5-meter and 10-meter
path lengths for absorbance measurements, and in a 1-meter path length
with built-in 365-nm excitation source for fluorescence measurements.
Each assembly has 560
µ
m inner-diameter Teflon
tubing with an internal
volume of 250
µ
l/meter, and will function with most liquids having a
refractive index >1.30.