beautypg.com

VEGA VEGADIF 55 4 … 20 mA_HART User Manual

Page 39

background image

3

D

oes the LED on the electronics module light briefly?

Y

es: existing pressure was accepted as position adjustment

N

o: Existing pressure for position adjustment was not accepted,

note the adjustment limits

Set lower pressure value:

1

R

equested pressure for lower pressure value (EMPTY PRES-

SURE)

is on the instrument

2

P

ush "-" key at least for 3 s

3

D

oes the LED on the electronics module light briefly?

Y

es: Existing pressure was saved as lower pressure value

(EMPTY PRESSURE)

and assigned to the lower level value

(EMPTY CALIB.)

N

o: Existing pressure was not saved as lower pressure value

(EMPTY PRESSURE). T

ake note of the adjustment limits

Set upper pressure value:

1

R

equested pressure for upper pressure value (FULL PRESSURE)

is acting on the instrument

2

P

ush "+" key at least for 3 s

3

D

oes the LED on the electronics module light briefly?

Y

es: Existing pressure was saved as upper pressure value (FULL

PRESSURE)

and assigned to the upper level value (FULL CALIB.)

N

o: Existing pressure was not saved as upper pressure value

(FULL PRESSURE). N

ote the adjustment limits.

I

f no local indication is connected, the following functions will be

possible via the three keys on the electronics module or outside on the
instrument:

l

P

osition adjustment (zero point correction)

l

S

et the max. pressure value and assign it to the max. flow value

l

I

nstrument reset, see chart in chapter "Operating elements"

Information:
T

he adjustment must be unlocked, see chapter "Lock/Unlock adjust-

ment".

B

y default, the instrument is set to mode Pressure. The mode can be

changed via the parameter "Measuring Mode", see chapter "Select
language and measuring mode
".

T

he "-" key is without function

T

he existing pressure must be within the nominal pressure limits of the

sensor, see specifications on the "Type label".

S

ee also "Quick setup menu" in chapter "Flow measurement".

Carry out position adjustment:

1

P

ressure on the instrument

Measuring Mode - Flow

VEGADIF

55 • 4 … 20

mA/HART

39

6 O

peration

31731

-EN

-081119