Averaging Level

Averaging Level

Averaging Level

Averaging Level

Averaging Level

Averaging level compensates for vibra-
tion or excessive air currents. During op-
eration, the balance continually takes
weight readings from the weighing cell.
Successive readings are then digitally pro-
cessed to achieve a stabilized display.
Use this submenu to specify how much
processing you need to obtain stable re-
sults.

NOTE

NOTE

NOTE

NOTE

NOTE: Averaging level does not affect
balance accuracy.

Select one of three averaging levels using
the adjacent table as a guide.

To view or change the averaging level:

1.

Access the AL submenu.

2.

Press

to display the current set-

ting.

3.

Press

to change the setting.

4.

Press

to accept the displayed

setting, when

released, AL will be

displayed again.

AVERAGING LEVEL

AVERAGING LEVEL

AVERAGING LEVEL

AVERAGING LEVEL

AVERAGING LEVEL

AL 0 Reduced stability, fastest

stabilization time.

AL 1

AL 1

AL 1

AL 1

AL 1 Normal stability, normal

Normal stability, normal

Normal stability, normal

Normal stability, normal

Normal stability, normal
stabilization time

stabilization time

stabilization time

stabilization time

stabilization time.....

AL 2 More stability, slow

stabilization time.