Ohaus CT Series ELECTRONIC BALANCE User Manual

Page 58

background image

17

Span Calibration

Span Calibration

Span Calibration

Span Calibration

Span Calibration

1.

Press and hold

until CAL is dis-

played, then release it. Balance will

display SPAN.

2.

Press

. When released, C 0g

will be displayed indicating that no

mass should be on the platform.

3.

Press

. The display will show

-C- followed by the value of the mass

which must be placed on the plat-

form.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: Do not disturb the balance when

-C- is displayed. Disturbances will result

in improper calibration.

4.

Place the required mass on the plat-

form and press

. The display will

show -C- while the balance recali-

brates. When the mass on the plat-

form is displayed along with the cur-

rent unit indicator, the balance is reca-

librated.