Parts Counting

Parts Counting

Parts Counting

Parts Counting

Parts Counting

Parts Counting must be enabled in the
Setup menu before using this procedure.
The balance will count parts based on the
weight of a reference sample, 5, 10, 20,
30, 40 or 50 parts. For optimum results,
the parts should be uniform in weight.

1.

Repeatedly press

until

PC

CON

is displayed.

2.

Place a container on the platform,

then press

. The balance dis-

plays

PC

ADD 5 which is the preset

reference quantity.

3.

To change the reference quantity,

repeatedly press

until the de-

sired quantity is displayed.

NOTE:

NOTE:

NOTE:

NOTE:

NOTE: If you need to exit or restart parts

counting at any time, press and hold

until

PC

CON is displayed, then release it.

Return to step 2 to restart, or press

for other weighing functions.

4.

Add the displayed number of parts to

the container, then press

. The

balance will display the current num-

ber of parts.