Ohaus CT Series ELECTRONIC BALANCE User Manual
Page 39

8
BY COMPUTER
DESCRIPTION
BALANCE RESPONSE
?
Print Display Units
The balance responds by sending the
current unit indicator and stability status in
the following format:
(Field)
Units Stability CR
(# of Chars.)
3
1
1
M
UNITS Button
This command has the same effect as
pressing the UNITS button on the front
panel. No data is output.
T
TARE Button
This command has the same effect as
pressing the TARE button on the front
panel. No data is output.
E
Extended UNITS
This command has the same effect as
Button
pressing and holding the UNITS button on
the front panel.
* L
Start Linearity
Starts Linearity Calibration procedure (see
Linearity Calibration in CT Series Manual).
* C
Start Calibration
Starts Span Calibration procedure (see
Span Calibration in CT Series Manual).
0D
Turn OFF 1 Second
Disables the 1 second delay the balance
Delay
waits before responding to a “P” command.
When disabled, the balance responds
immediately.
1D
Turn ON 1 Second
Enables a 1 second delay in the balance’s
Delay
response to a “P” command. This is
provided for computers requiring such a
delay.
* 0S
Turn OFF
Disables the Stable Data Only printing
"Stable Data Only"
feature. When disabled, the balance will
send display data in response to a “P”
command, regardless of the stability status.
* 1S
Turn ON
Enables the Stable Data Only printing
"Stable Data Only"
feature. When the feature is enabled, the
balance will only send stable display data in
response to the “P” command.
* Does not apply to Legal For Trade models.