beautypg.com

Ohaus CT Series ELECTRONIC BALANCE User Manual

Page 39

background image

8

BY COMPUTER

DESCRIPTION

BALANCE RESPONSE

?

Print Display Units

The balance responds by sending the
current unit indicator and stability status in
the following format:

(Field)

Units Stability CR

(# of Chars.)

3

1

1

M

UNITS Button

This command has the same effect as
pressing the UNITS button on the front
panel. No data is output.

T

TARE Button

This command has the same effect as
pressing the TARE button on the front
panel. No data is output.

E

Extended UNITS

This command has the same effect as

Button

pressing and holding the UNITS button on

the front panel.

* L

Start Linearity

Starts Linearity Calibration procedure (see
Linearity Calibration in CT Series Manual).

* C

Start Calibration

Starts Span Calibration procedure (see
Span Calibration in CT Series Manual).

0D

Turn OFF 1 Second

Disables the 1 second delay the balance

Delay

waits before responding to a “P” command.
When disabled, the balance responds
immediately.

1D

Turn ON 1 Second

Enables a 1 second delay in the balance’s

Delay

response to a “P” command. This is
provided for computers requiring such a
delay.

* 0S

Turn OFF

Disables the Stable Data Only printing

"Stable Data Only"

feature. When disabled, the balance will
send display data in response to a “P”
command, regardless of the stability status.

* 1S

Turn ON

Enables the Stable Data Only printing

"Stable Data Only"

feature. When the feature is enabled, the
balance will only send stable display data in
response to the “P” command.

* Does not apply to Legal For Trade models.

COMMAND SENT