Rs232 error messages, Rs-232 error messages – Ohaus CT Series ELECTRONIC BALANCE User Manual
Page 40

9
BY COMPUTER
DESCRIPTION
BALANCE RESPONSE
nnA
Set Auto Print
Set the Auto Print Interval to “nn” seconds
Interval
(nn = 0 to 99). Setting the interval to 00
disables the Auto Print function.
* 0Z
Turn OFF Auto Zero Disables the Auto Zero Tracking feature
Tracking
(see Setup Menu in CT Series Manual).
* 1Z
Turn ON Auto Zero
Enables the Auto Zero Tracking feature
Tracking
(see Setup Menu in CT Series Manual).
** V
Print PROM
Causes the balance to respond with the
Revision Level
revision level of the PROM installed in the
balance. Output is in the format: “CT V_._”.
** N
Disable Automatic
Permits you to disable the automatic printing
Error Message
of error messages output by the balance.
Printing
This command is reset on power-off.
**
Revert to Factory
Sets all balance menu selections to the
Default Settings
following factory defaults:
All available units enabled
Default units = Grams
* Auto-Zero Tracking is ON
High Speed Display Update is OFF
* Custom Unit = Carats
* Custom Unit Conversion Factor is set to:
1.0000, LSD = 1, Exponent = 0
“Stable Data Only” printing is ON
Auto Print feature is OFF
* Does not apply to Legal For Trade models.
** Available only on balances with version 2.0 software or later.
RS-232 Error Messages
RS-232 Error Messages
RS-232 Error Messages
RS-232 Error Messages
RS-232 Error Messages
If the balance is in an error condition when a request for output occurs, the current error
message number (Err0 - Err6) will be sent in the following format:
(Field)
Error # CR LF
(# of Chars.)
5
1 1