Calibration menu – Ohaus CT Series ELECTRONIC BALANCE User Manual
Page 57
16
CALIBRATION MENU
CALIBRATION MENU
CALIBRATION MENU
CALIBRATION MENU
CALIBRATION MENU
Portable
Advanced
balances can be calibrated in two ways: Span calibration or
Linearity calibration. Span calibration resets the balance’s weighing range using two
points: zero and a calibrated mass value at or near the balance’s capacity. Linearity
calibration minimizes deviation between actual and displayed weights within the
balance’s weighing range using three points: zero, a calibrated mass value within the
balances weighing range, and a value at or near the balance’s specified capacity. The
illustration below indicates the sequence in which submenus appear on the Calibration
menu.
CAL
SPAN
LINEARITY
END
Calibration Menu Protection
Calibration Menu Protection
Calibration Menu Protection
Calibration Menu Protection
Calibration Menu Protection
The Calibration menu may be locked out
to prevent unauthorized access. To lock
out calibration menu, refer to the section
titled Calibration Lock-Out Protection.
Calibration Masses
Calibration Masses
Calibration Masses
Calibration Masses
Calibration Masses
Before beginning calibration, make sure
calibrated masses are on hand. If you
begin calibration and realize masses are
not available, either turn the balance off, or
go through the procedure without masses.
The balance will use previously stored
calibration data. Calibration should be
performed as necessary to ensure accu-
rate weighing. Masses required to per-
form the procedures are listed in the adja-
cent table.
Model Span Linearity
Model Span Linearity
Model Span Linearity
Model Span Linearity
Model Span Linearity
Masses Masses
Masses Masses
Masses Masses
Masses Masses
Masses Masses
CT10 10g 5g, 10g
CT200 200g
100g, 200g
CT600 300g
(2) 300g
CT1200 1000g 500g, 1000g
CT6000 2000g
(3) 2000g
1000g
1000g
Required Calibration Masses
Required Calibration Masses
Required Calibration Masses
Required Calibration Masses
Required Calibration Masses