beautypg.com

Ohaus CT Series ELECTRONIC BALANCE User Manual

Page 60

background image

19

6.

Press

. The display will show

-C- momentarily, then C followed by

the next mass to be placed on the

platform. Do not disturb the balance

when -C- is displayed. Disturbances

will result in improper calibration.

7.

Place the required mass on the plat-

form, then press

. The display

will show -C- while the balance reca-

librates. When the mass on the plat-

form is displayed along with the cur-

rent unit indicator, the balance is reca-

librated.

End

End

End

End

End

If you have entered the Calibration menu
and do not wish to calibrate the balance,
use END to return to normal weighing
operations.

1.

Repeately press

until END is

displayed.

2.

Press

. When released, the bal-

ance will returned to normal weighing

operations.