beautypg.com

M&C TechGroup PMA 20 Operator's manual User Manual

Page 20

background image

20

Gas sampling and gas conditioning technology

9-3.5.1-ME

Connect the hose end of the instrument air or check gas bottle connection to the gas inlet

of the analyser.

Open the pressure reducer valve slowly, to avoid pressure peaks.

Adjust the flow rate to 50 Nl/hr at the flow meter.

N O T E !



Always calibrate at the flow rate that is adjusted for the measurement
too.

Wait approximately 20

– 30 sec. until the indication has stabilised.

If necessary adjust span accurately according to the check gas concentration with a screw

driver at the span potentiometer in the front panel. In case of air e.g. to 20,9 % O

2

.

Check output signals at 20,9% O

2

:



Shut off pressure reducer valve and bottle- resp. Instrument air valve resp. integrated

sample gas pump.

Disconnect hose connection at the analyser.


Determination of the output signal:

(S

e

– S

np

) V resp. mA x gas concentration Vol % O

2

upper range value Vol % O

2


Se = Final value, signal output
Snp = Zero, signal output

Shut off pressure reducer output valve and bottle valve. Druckreglerausgangsventil und

Flaschenventil schließen. Disconnect hose connection at the analyser.


The span calibration is finished.

N O T E !


If during the span calibration great variations have to be compensated
(>2% O

2

) at the potentiometers, a second zero and span calibration is

reasonable.

A T T E N T I O N

After finishing the calibration turn measuring selection switch to the

desired measuring range.
The mA-output signal is dependant on the measuring range!





Output signal

Measurement
range 100 % O

2

Measurement
range 30 % O

2

0-1 V

0,209 V

0,697 V

0-20 mA

4,18 mA

13,93 mA

4-20 mA

7,34 mA

15,15 mA

+ S

np